• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON MEDICAL (SINGAPORE) BD ECLIPSE¿ NEEDLE 18 G X 1 1/2 IN.; HYPODERMIC SINGLE LUMEN NEEDLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON MEDICAL (SINGAPORE) BD ECLIPSE¿ NEEDLE 18 G X 1 1/2 IN.; HYPODERMIC SINGLE LUMEN NEEDLE Back to Search Results
Model Number 305766
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/18/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The date received by manufacturer has been used for this field.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that prior to use of a bd eclipse¿ needle 18 g x 1 1/2 in., that foreign matter was observed.The following information was provided by the initial reporter: staff members just noticed a black marking on the hub of an 18 ga, 1 ½ ¿ needle.
 
Event or Problem Description
It was reported that prior to use of a bd eclipse¿ needle 18 g x 1 1/2 in., that foreign matter was observed.The following information was provided by the initial reporter: staff members just noticed a black marking on the hub of an 18 ga, 1 ½ ¿ needle.
 
Additional Manufacturer Narrative
H.6.Investigation: one photo was received by our quality team for evaluation.From the photo, black foreign matter was observed inside the eclipse hub.A review of the internal manufacturing device records and raw material history files for the reported lot number was performed and no recorded quality problems or rejections to this incident were found.As the actual sample was not returned to determine the foreign matter type.Therefore, the root cause could not be established.H3 other text : see h.10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD ECLIPSE¿ NEEDLE 18 G X 1 1/2 IN.
Common Device Name
HYPODERMIC SINGLE LUMEN NEEDLE
Manufacturer (Section D)
BECTON DICKINSON MEDICAL (SINGAPORE)
30 tuas avenue 2
singapore
Manufacturer (Section G)
BECTON DICKINSON MEDICAL (SINGAPORE)
30 tuas avenue 2
singapore
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key13656375
Report Number8041187-2022-00107
Device Sequence Number12106336
Product Code FMI
UDI-Device Identifier30382903057666
UDI-Public30382903057666
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K161170
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date09/30/2026
Device Model Number305766
Device Catalogue Number305766
Device Lot Number1272165
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/18/2022
Supplement Date Received by Manufacturer04/05/2022
Initial Report FDA Received Date03/02/2022
Supplement Report FDA Received Date04/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/29/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-