• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICRO THERAPEUTICS, INC. DBA EV3 AXIUM PRIME BRPL 3D; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICRO THERAPEUTICS, INC. DBA EV3 AXIUM PRIME BRPL 3D; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION Back to Search Results
Model Number APB-1-2-3D-ES
Medical Device Problem Codes Separation Failure (2547); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/17/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information that an axium coil prematurely detached.The patient was being treated for a ruptured fusiform aneurysm in the middle carotid artery (mca).The max diameter was 3mm and the neck diameter was 4mm.Patient blood flow was normal and vessel tortuosity was moderate.The devices were prepared according to the instructions for use (ifu).  a coil form another company was deployed.Then a second coil which was from medtronic was prematurely detached.Some coil part came in to the parent artery.Physician then chose to deploy an atlas stent.The coil remains in patient.The coil was implanted at the intended location.No surgical or medical intervention is required.The pushwire was not bent or broken.The physician repositioned the coil 2 times.The physician attempted to detach the coil.They tried 2 times with the instant detacher.They tried 1 time using the manual method.1 instant detacher was used.The physician didn't not rotate the delivery pusher during the procedure.Continuous flush was administered during the procedure.The patient is fine.No symptoms were reported.  ancillary devices: neuron guide catheter, sl 10 microcatheter, synchro guidewire.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AXIUM PRIME BRPL 3D
Common Device Name
DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer (Section G)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer Contact
glen belmer
9775 toledo way
irvine, CA 92618
6122713209
MDR Report Key13656388
Report Number2029214-2022-00338
Device Sequence Number20177327
Product Code KRD
UDI-Device Identifier00847536031814
UDI-Public00847536031814
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K151447
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/09/2024
Device Model NumberAPB-1-2-3D-ES
Device Catalogue NumberAPB-1-2-3D-ES
Device Lot NumberB203442
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/17/2022
Initial Report FDA Received Date03/02/2022
Date Device Manufactured05/10/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age50 YR
Patient SexFemale
Patient Weight67 KG
-
-