| Catalog Number |
FND-020-01/B |
| Medical Device Problem Codes |
Adverse Event Without Identified Device or Use Problem (2993); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Codes |
Hemorrhage/Blood Loss/Bleeding (1888); Low Blood Pressure/ Hypotension (1914); Perforation (2001)
|
| Date of Event |
02/22/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The suspect device is not expected to return for evaluation.A follow up will be submitted when the evaluation is complete.
|
| |
|
Event or Problem Description
|
|
The account alleges that during an a-fib procedure, while under general anesthesia, during the mapping phase, the anesthesiologists noticed a depreciation of the vital parameters, such as blood pressure drop.The transesophageal echo showed fluid in the pericardium, thus a pericardiocentesis was performed for draining the blood.As the fluid drained, it was concluded that open chest surgery was needed for the bleeding to be stopped.The patient was transferred to the or and the surgeons found a perforation in the left atrial appendage (laa).The patient was transferred to the icu post-op.No additional patient consequence to report.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect device was not returned for evaluation.The complaint could not be confirmed.The root cause could not be determined.A search of the complaint database was performed and no similar complaints for this lot number were found.The device history record was reviewed, and no exception documents were found.
|
| |
|