• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD ICON SERIES CPAP; BZD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD ICON SERIES CPAP; BZD Back to Search Results
Model Number ICONAAN
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/28/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The complaint icon cpap is currently is en-route to fisher and paykel healthcare (f&p) for evaluation.We will provide a follow-up report upon completion of our investigation.
 
Event or Problem Description
A customer in (b)(6), reported that the power cord of an icon cpap humidifier was damaged and had exposed copper wires.There was no patient consequence.
 
Event or Problem Description
A customer in ohio reported that the power cord of an icon cpap humidifier was damaged and had exposed copper wires.There was no patient consequence.
 
Additional Manufacturer Narrative
(b)(4).Method: the complaint icon auto cpap humidifier was not returned to fisher & paykel healthcare (f&p) for evaluation.Our investigation is based on the photography and description of events provided by the customer and our knowledge of the product.Results: visual inspection of the photography provided by the customer revealed that the power cord is damaged.Conclusion: based on the investigation conducted, the power cord may have been subjected to excessive force, causing the reported damage.During initial assembly of the icon auto cpap, all power cords are visually inspected for damage.Once the assembly process is completed, all icon auto cpap devices are visually inspected again before release for distribution.This suggests the damage occured after it had been distributed.The icon auto cpap is designed to the electrical safety standards, ul60601-1 and as/nzs 3200.1.The materials used in the thermoplastic components of the mains connector and the cases are flame retardant according to ul 60601-1 and as/nzs3200.1.Our user instructions that accompany the icon auto cpap humidifier state the following: - "only operate if the device, power cord and plug are dry and in good working order." - "do not operate the device, water chamber or breathing tube if it is dropped, damaged or not working as intended.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ICON SERIES CPAP
Common Device Name
BZD
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
faranak gomarooni
173 technology drive
suite 100
irvine, CA 92618
9494534000
MDR Report Key13657125
Report Number9611451-2022-00207
Device Sequence Number4984851
Product Code BZD
UDI-Device Identifier09420012425027
UDI-Public(01)09420012425027(10)2100019692(11)160120
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K094040
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 02/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberICONAAN
Device Catalogue NumberICONAAN
Device Lot Number2100019692
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/24/2022
Supplement Date Received by Manufacturer03/24/2022
Initial Report FDA Received Date03/02/2022
Supplement Report FDA Received Date04/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/20/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-