• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION MUSTANG; CATHETER, BILIARY, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION MUSTANG; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number 24674
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/11/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon rupture occurred.Th patient presented with thrombosis.The target lesion was located in the iliac vein.A 12.0 x 80, 135cm mustang balloon catheter was advanced for dilation.However, during inflation, the balloon burst.The procedure was completed with another of the same device.There were no patient complications and the patient status was stable.
 
Event or Problem Description
It was reported that balloon rupture occurred.The target lesion was located in the iliac vein.A 12.0 x 80, 135cm mustang balloon catheter was advanced for dilation.However, during inflation, the balloon burst.The procedure was completed with another of the same device.There were no patient complicaitons and the patient was stable.It was further reported that the balloon ruptured during first inflation at less working pressure for few seconds.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: a mustang was returned analysis.A visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.A longitudinal tear was identified in the balloon material.The tear measured approximately 85mm in length and extended from a position 10mm proximal of the proximal marker band to 11mm distal of the distal marker band.An examination of the balloon material and marker bands identified no issues which could potentially have contributed to this complaint.A visual examination observed no damage to the tip of the device.A visual and tactile examination found no kinks or damage to the shaft of the device.Both marker bands were undamaged and present on the shaft of the device.
 
Event or Problem Description
It was reported that balloon rupture occurred.The target lesion was located in the iliac vein.A 12.0 x 80, 135cm mustang balloon catheter was advanced for dilation.However, during inflation, the balloon burst.The procedure was completed with another of the same device.There were no patient complications and the patient was stable.It was further reported that the balloon ruptured during first inflation at less working pressure for few seconds.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MUSTANG
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key13657615
Report Number2134265-2022-02382
Device Sequence Number19318958
Product Code FGE
UDI-Device Identifier08714729795285
UDI-Public08714729795285
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K141521
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/29/2024
Device Model Number24674
Device Catalogue Number24674
Device Lot Number0027584374
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/24/2022
Supplement Date Received by Manufacturer03/02/2022
07/21/2022
Initial Report FDA Received Date03/03/2022
Supplement Report FDA Received Date03/09/2022
07/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/30/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age55 YR
Patient SexMale
Patient Weight70 KG
-
-