| Model Number |
NNR-0F |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/17/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A cervical kerrision broke during the procedure while in use.All pieces wer retrieved by the surgeon.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the lot number, we have checked the dhr: no findings we will examine the device as soon as possible.
|
| |
|
Event or Problem Description
|
|
A cervical kerrision broke during the procedure while in use.All pieces wer retrieved by the surgeon.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the lot number, we have checked the dhr: no findings.The foot of the punch broke off, but the broken piece wasn't returned.Microscope photos show serious deformations of the cutting edge of the slider.Based on these findings, we assume that the breakage was caused by overloading due to cutting too hard material.(see enclosed pictures) the punch is made of hardened steel and is used for punching bone structures thinned out by milling.When used properly, it is not possible for lippage of this type to occur on the cutting edge.A material or manufacturing defect can be ruled out due to the homogeneous structure of the fracture surface.
|
| |
|