MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PFC SIGMA MOD TIB TRAY CEM SZ2; SIGMA KNEE PRIMARY : KNEE TIBIAL TRAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PFC SIGMA MOD TIB TRAY CEM SZ2; SIGMA KNEE PRIMARY : KNEE TIBIAL TRAY Back to Search Results
Model Number 86-6022
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Fall (1848); Foreign Body Reaction (1868); Pain (1994); Osteolysis (2377); Swelling/ Edema (4577); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 12/14/2021
Type of Reportable Event Serious Injury
Event or Problem Description
Subject id: (b)(6).Study no: (b)(6).Clinical notification received for revision to address stress reaction.Date of implantation: (b)(6) 2003.Date of revision: (b)(6) 2021 (right knee).Treatment: revision of the tibial insert.Clinic visit (b)(6) 2021: patient presents with right knee pain, swelling, warmth, and crepitus after several falls 20 years status post bilateral knee implantation.Radiographic images identify an anteriorly distracted tibial insert and an area of asymptomatic tibial osteolysis.Plan is made to revise the insert, as the tibial and femoral components are well-fixed and intact.Dor: (b)(6) 2021: patient received a right knee insert revision to treat pain and limited range of motion secondary to displacement of the tibial insert.Upon entering the joint, the surgeon identified black stained synovium and performed a complete synovectomy.The insert was displaced anteriorly.The surgeon notes there was no wear or problems with the locking mechanism on the tray.The tray was well-fixed and retained.The issue was resolved by replacing the old insert with a new insert that was one size larger.The insert snapped into position and a rom check confirmed it was not displaceable.The procedure was completed without complications.Doi: unknown.Dor: (b)(6) 2021; right knee.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PFC SIGMA MOD TIB TRAY CEM SZ2
Common Device Name
SIGMA KNEE PRIMARY : KNEE TIBIAL TRAY
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS 1818910
700 orthopaedic dr
warsaw IN 46581 0988
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key13658525
Report Number1818910-2022-03960
Device Sequence Number1869693
Product Code JWH
UDI-Device Identifier10603295216773
UDI-Public10603295216773
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K884796
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2003
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Study,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number86-6022
Device Catalogue Number866022
Device Lot Number1058701
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/15/2022
Supplement Date Received by Manufacturer03/10/2022
Initial Report FDA Received Date03/03/2022
Supplement Report FDA Received Date03/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/17/2002
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PFC SIGMA STAB INS SZ2 10MM; PFC*SIGMA C/S NPOR FE RT SZ2.5; SIGMA PATELLA OVAL DOME 3 PEGGED 35MM
Outcome Attributed to Adverse Event Required Intervention;
Patient Age79 YR
Patient SexFemale
Patient Weight88 KG
-
-