• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC NESTER PLATINUM EMBOLIZATION MICROCOIL; KRD DEVICE, EMBOLIZATION, VASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC NESTER PLATINUM EMBOLIZATION MICROCOIL; KRD DEVICE, EMBOLIZATION, VASCULAR Back to Search Results
Model Number N/A
Medical Device Problem Codes Material Separation (1562); Inadequacy of Device Shape and/or Size (1583)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/21/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported following deployment of nester platinum embolization microcoil to treat a type 3 endoleak, the coil remained stretched instead of coiling up.The physician then attempted to remove the coil using a snare, but it "fell apart." the patient had to be opened to remove the device.Aortic stents were also removed and replaced, presumably to treat the type 3 endoleak.The physician later clarified that the coil was deployed and "not faulty," but separated during snaring.He also stated "everything went fine." no other adverse effects were reported.
 
Additional Manufacturer Narrative
This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported the patient's hospitalization was prolonged due to this occurrence.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unchanged, unknown, or unavailable.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Correction: d8 investigation ¿ evaluation presbyterian hosp (united states) informed cook on (b)(6) 2022 of an incident that occurred on the same date involving an mwce-18-14-6-nester-01 (nester platinum embolization microcoil) from lot 14376968.When the user deployed the coil, it remained elongated; therefore, removal with a snare was attempted.At this time, the device ¿fell apart¿, so open surgery was performed to remove the device.No other adverse events were reported due to this occurrence.Reviews of the complaint history, device history record (dhr), quality control, and instructions for use (ifu) were conducted during the investigation.The complaint device was not returned; therefore, no physical examinations could be performed.However, a document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.A review of the dhr for the reported complaint device lot 14376968 revealed no relevant non-conformances.A database search did not identify any other events associated with the reported device lot.Based on the available information, cook has concluded that the device was manufactured to specification and that there is no evidence suggesting nonconforming product exists either in house or in field.Cook also reviewed product labeling.The current instructions for use [t_nec2_rev0] does not have wording about this specific failure.Based on the available information, no device return, and the results of the investigation, cook has concluded that component failure unrelated to manufacturing or design deficiencies contributed to this incident.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints.Per the risk assessment no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NESTER PLATINUM EMBOLIZATION MICROCOIL
Common Device Name
KRD DEVICE, EMBOLIZATION, VASCULAR
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key13659199
Report Number1820334-2022-00329
Device Sequence Number5046300
Product Code KRD
UDI-Device Identifier00827002473294
UDI-Public(01)00827002473294(17)261230(10)14376968
Combination Product (Y/N)N
Initial Reporter StateNM
Initial Reporter CountryUS
PMA/510(K) Number
K153778
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue NumberMWCE-18-14-6-NESTER-01
Device Lot Number14376968
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/21/2022
Supplement Date Received by Manufacturer04/12/2022
11/23/2022
Initial Report FDA Received Date03/03/2022
Supplement Report FDA Received Date04/13/2022
12/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/30/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age57 YR
Patient SexMale
-
-