| Model Number |
N/A |
| Medical Device Problem Codes |
Positioning Failure (1158); Off-Label Use (1494)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/08/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a total knee arthroplasty, the bearing would not mate with corresponding femur implant.It was believed that this device was compatible with the corresponding implants.When the surgeon could not get a proper fit, they sawed off part of the post on this device to adjust the fit.There was a surgical delay of 30 minutes and it's been reported that the patient is waiting for the correct implant, indicating a revision is likely to occur.Attempts have been made and no further information has been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Customer has indicated that the product will not be returned to zimmer biomet for investigation, as the device remains implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Report source: (b)(6).Concomitant products: cp113123, vngd ti fem ssk 60mm lt, lot 934750; 141205, biomet tibial locking bar, lot 602980.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.No product or pictures were provided, as the device remains implanted.Review of the device history records identified no related deviations or anomalies during manufacturing.The reported products were reviewed for compatibility and it was determined that items 183868 (vngd ssk psc tib brg 18x63/67 7 x 18) and 185184 (vngd ssk 360 psc tib brg 18x59 18x59mm) are not compatible with item cp113123 (vngd ti fem ssk 60mm lt).The root cause of the reported issue is attributed to the off-label use of devices, as vanguard psc bearings are not compatible with the ssk femur used.It is unknown if this off-label use caused or contributed to the additional issue with the bending of the locking bar noted during removal.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
It was reported that during a total knee arthroplasty, the bearing would not mate with corresponding femur implant.It was believed that this device was compatible with the corresponding implants.When the surgeon could not get a proper fit, they sawed off part of the post on this device to adjust the fit.There was a surgical delay of 30 minutes and it's been reported that the patient is waiting for the correct implant, indicating a revision is likely to occur.Further follow-ups state that the patient will not undergo a revision, as they are satisfied with the knee at this time.No harm or impact has occurred.No additional information is available.
|
| |
|