• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAYER MEDICAL CARE, INC. STELLANT DUAL CT INJECTOR; CT INJECTION SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAYER MEDICAL CARE, INC. STELLANT DUAL CT INJECTOR; CT INJECTION SYSTEM Back to Search Results
Model Number 85631829
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/16/2022
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported the following: as the staff was preparing the stellant flex system (serial number (b)(4)) for a procedure, the injector head disengaged from the ocs iii system (ocs iii 0214/17677).No injury or adverse event was reported.
 
Additional Manufacturer Narrative
A bayer service representative responded to the site and replaced the pivot knuckle assembly which restored the injection system to normal operation.Bayer product analysis visually examined a photograph provided for investigation and confirmed a sheared bushing screw within the pivot knuckle assembly.The sheared bushing screw allowed the pivot knuckle to release and the injector head to disengage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STELLANT DUAL CT INJECTOR
Common Device Name
CT INJECTION SYSTEM
Manufacturer (Section D)
BAYER MEDICAL CARE, INC.
1 bayer drive
indianola PA 15051
Manufacturer (Section G)
BAYER MEDICAL CARE, INC.
1 bayer drive
indianola PA 15051
Manufacturer Contact
susan sherwin
1 bayer drive
indianola, PA 15051
7249408678
MDR Report Key13660514
Report Number2520313-2022-00009
Device Sequence Number1945198
Product Code DXT
UDI-Device Identifier00616258020792
UDI-Public(01)00616258020792(11)120401
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K192370
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number85631829
Device Catalogue NumberSCT FLEX
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/16/2022
Initial Report FDA Received Date03/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/14/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
OCS III 0214/17677
Patient SexUnknown
-
-