| Model Number |
BA25-100/I16-40 |
| Medical Device Problem Codes |
Material Puncture/Hole (1504); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Aneurysm (1708); Failure of Implant (1924); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
02/21/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The patient was initially treated for an abdominal aortic aneurysm (aaa) with implant of an afx bifurcated stent graft and an afx vela suprarenal.Approximately six (6) and a half years post initial procedure, at routine follow-up a type ii endoleak (non-device related) from patent lumbers and inferior mesenteric artery was identified with suspect type iiib endoleak.Diagnostic angiogram is planned but has not been performed.The patient was reported to be asymptomatic.
|
| |
|
Additional Manufacturer Narrative
|
|
The devices involved in this event will not be returned for evaluation and remain implanted in the patient.Patient medical records and imaging studies will be requested for further evaluation by a clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.Device iteration is afx with duraply.Device remains implanted.
|
| |
|
Event or Problem Description
|
|
The patient was initially treated for an abdominal aortic aneurysm (aaa) with implant of an afx bifurcated stent graft and an afx vela suprarenal.Approximately six (6) and a half years post initial procedure, at routine follow-up a type ii endoleak (non-device related) from patent lumbers and inferior mesenteric artery was identified with suspect type iiib endoleak.Diagnostic angiogram is planned but has not been performed.The patient was reported to be asymptomatic.Subsequent to the initial report, additional information was received reporting that angiogram was performed on (b)(6) 2022 and confirmed that there was only a type ii endoleak which was embolized.The afx bifurcated stent graft was intact with no issues.Patient is doing well.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good faith efforts to retrieve a reported adverse event/incident related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.An evaluation of the device could not be completed.The device was not returned to endologix for evaluation because it remains implanted.A clinical evaluation of the adverse event/incident was completed.An examination of medical records and/or medical imaging received by endologix shows the type ii endoleak is confirmed.This is consistent with the reported adverse event/incident.The most likely causation for the type ii endoleak is anatomy related.The procedure related harms for this complaint could not be determined.The contributing factor for the reported event is most likely the patent lumbar artery mentioned on (b)(6) 2015 operative notes.The final patient status was reported being stable post the embolization procedure.Type ii endoleaks are anatomy-related events and are not caused or contributed by the device.A complaint is not warranted since there is no alleged deficiency related to the identity, quality, durability, reliability, safety, effectiveness, or performance of the device.Therefore, this is no longer a reportable event.Device iteration is afx with duraply.B5: describe event or problem - updated.G4: awareness date ¿ updated.H6: patient code - removed 1708; 1924.H6: health effect - impact code - removed 4614.H6: medical device problem codes - removed 1504.H6: investigation finding codes - removed 3233.H6: investigation conclusion codes - removed 11.
|
| |
|