Model Number 0046180
Medical Device Problem Code Unclear Information (4052)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/03/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
 
Event or Problem Description
It was reported that the nasogastric tubes did not have numbers on them and therefore could not be accurately measured and communicated to the next care provider with certainty.Nasogastric tube was advanced more and confirmed with air blow, but still there was no numbers on the nasogastric tube to confirm placement while patient was in the hospital.The patient went for x-ray to confirm placement and nasogastric tubes was not in correct spot by over 10cm.
 
Event or Problem Description
It was reported that the nasogastric tubes did not have numbers on them and therefore could not be accurately measured and communicated to the next care provider with certainty.Nasogastric tube was advanced more and confirmed with air blow, but still there was no numbers on the nasogastric tube to confirm placement while patient was in the hospital.The patient went for x-ray to confirm placement and nasogastric tubes was not in correct spot by over 10cm.
 
Additional Manufacturer Narrative
The reported event was inconclusive because no sample was returned.A potential root cause for this failure could be "the ribbon jams in the machine/ end of ribbon".It is unknown whether the device had met relevant specifications.The product was used for treatment purposes.It was unknown whether the product had caused the reported failure.A review of the device history record did not show any problem or conditions that would have contributed to the reported issue.Based on the results of the investigation, no additional actions are needed.The device was not returned for evaluation.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: instructions for statlock® nasogastric stabilization device (when included): remove oil and moisturizer from targeted skin area.Apply skin prep to targeted statlock® stabilization device area for skin protection and enhanced pad adherence.Allow to dry completely.Caution: federal (u.S.A.) laws restricts this device to sale by or on the order of a physician.Sterile unless package is opened or damaged, except for any individually packaged components within the pouch which are not labeled as sterile.These components are not terminally sterilized.Single use.Do not resterilize.Statlock® stabilization device does not contain natural rubber latex.Note: after use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practice and applicable local, state and federal laws and regulations.H11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.H3 other text: the device was not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® NASOGASTRIC SUMP TUBE WITH PREVENT® ANTI-REFLUX FILTER
Common Device Name
NASOGASTRIC SUMP TUBE WITH PREVENT
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
juan velez
8195 industrial blvd
covington 30014
7707846100
MDR Report Key13660895
Report Number1018233-2022-01017
Device Sequence Number14684995
Product Code FEG
UDI-Device Identifier00801741052347
UDI-Public(01)00801741052347
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K960176
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2024
Device Model Number0046180
Device Catalogue Number0046180
Device Lot NumberNGDQ1394
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/07/2022
Supplement Date Received by Manufacturer07/06/2022
Initial Report FDA Received Date03/03/2022
Supplement Report FDA Received Date07/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/07/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-