Model Number 108-ET-6540C
Medical Device Problem Codes Fracture (1260); Mechanical Problem (1384)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 02/10/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Though revision surgery occurred, pathology department would not release device, and the devices were not returned.Device is not available for evaluation.Additionally, no event radiographs of the device were provided for analysis.Photo of the fractured screw confirmed the event.Unknown factors include: patient activity at the time or prior to the event, patient bone quality, the degree of spinal instability, the longevity the degree of post lateral pseudarthrosis, patient compliance with post-operative care instructions or if the patient sustained a fall/impact of any sort (no trauma was reported.).Root cause or specific failure mode cannot be determined.
 
Event or Problem Description
As part of a retrospective review spanning 2019-2021 given by an orthopedic medical group, it was reported that a patient received a plf with bilateral posterior fixation from l5-s1.Revision surgery was performed on (b)(6) 2022.Photograph was provided of one of the s1 bone screws that fractured.No initial surgery information, patient information or event information were provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OVERWATCH
Common Device Name
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
Manufacturer (Section D)
SPINAL ELEMENTS, INC
3115 melrose drive, suite 200
carlsbad CA 92010
Manufacturer Contact
peter perhach
3115 melrose drive, suite 200
carlsbad, CA 92010
7606071830
MDR Report Key13660969
Report Number3004893332-2022-00005
Device Sequence Number5009495
Product Code NKB
UDI-Device Identifier00840916131852
UDI-Public(01)00840916131852(10)201935(11)201203
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K161842
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number108-ET-6540C
Device Catalogue Number108-ET-6540C
Device Lot Number201935
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/10/2022
Initial Report FDA Received Date03/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/03/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-