Model Number RT380 |
Device Problems
Failure of Device to Self-Test (2937); Gas/Air Leak (2946)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/01/2022 |
Event Type
malfunction
|
Event Description
|
A distributor in (b)(4) reported on behalf of a healthcare facility, via a fisher & paykel healthcare (f&p) field representative, that an rt380 adult dual-heated evaqua2 breathing circuits did not pass the leak test.There was no patient involvement.
|
|
Manufacturer Narrative
|
(b)(4).The complaint rt380 adult dual-heated evaqua2 breathing circuit is currently en route to fisher & paykel healthcare (f&p) for evaluation.We will provide a follow up report upon completion of investigation.
|
|
Event Description
|
A distributor in japan reported on behalf of a healthcare facility, via a fisher & paykel healthcare (f&p) field representative, that an rt380 adult dual-heated evaqua2 breathing circuit did not pass the leak test.There was no patient involvement.
|
|
Manufacturer Narrative
|
(b)(4).Method: the complaint rt380 adult dual heated evaqua2 breathing circuit kit was returned to fisher & paykel healthcare (f&p) in new zealand for evaluation where it was visually inspected and pressure tested.Results: visual inspection of the returned circuit identified no damage or defects.Pressure testing of the circuit tubing revealed the circuit was within specification.Visual inspection of the chamber from the kit identified a tear in the water feedset tube near the waterbag spike.Conclusion: the reported event was likely due to the tear observed in the water feedset tubing.We are unable to determine the cause of the water feedset damage.Every mr290v chamber is pressure tested following the manufacturing process to check for any leaks present in the chamber due to cracks and other causes.Any chamber that fails this test is rejected.Also, the pressure test is followed by a visual inspection of each chamber.In addition, pull testing is performed periodically to verify the strength of the water feedset tubing.The subject mr290v chamber would have met the required specification at the time of production.All rt380 adult evaqua2 breathing circuits are visually inspected and pressure and flow tested during production, and those that fail are rejected.The subject breathing circuit would have met the required specifications at the time of production.Our user instructions that accompany the rt380 adult evaqua2 breathing circuit show in pictorial format the correct set-up of the circuit and also states the following: "check all connections are tight before use." "perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient." "ensure appropriate ventilator or flow source alarms are set before connecting breathing set to patient.".
|
|
Search Alerts/Recalls
|