| Model Number |
JK100 |
| Medical Device Problem Code |
Structural Problem (2506)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Manufacturing site evaluation: investigation on-going.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported to aesculap ag that a filter retention plate (part # jk100) was used during a procedure performed on an unknown date.According to the complainant, the filter holders spring off.The sample was produced in 2006 and repaired by moritz in 2017 [mm 97/18].On the basis of the gasket, it was determined that yard goods were installed.The complaint device has been returned to the manufacturer for evaluation.There was no described patient harm.Although requested, additional information has not been made available.The malfunction is filed under aag reference (b)(4).Involved components: jk486 - full-size lid w/retention plate blue - lot unknown.
|
| |
|
Event or Problem Description
|
|
Correction: patient harm changed from "na to significant surgery delay".There was a significant surgery delay.Additional information was not provided nor available.Additional information has been requested but not yet received as of this report.Additional patient information is not available.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction / additional information: b5: patient harm changed from "na to significant surgery delay".H3: device evaluated by manufacturer.H6: codes.Investigation results: visual investigation: the filter holders show no visible damages.One filter holding pin of the retention plate is deformed.The investigation was carried out visually with the digital camera.We carried out a visual investigation of the products and noted, that one filter holder pin of the retention plate is deformed.Batch history review: due to the fact that no lot number was provided, a review of the device history records for the complained device is not possible.The review of risk assessment revealed that the overall risk level (severity 1(5) x probability of occurrence 3(5)) according to din en iso 14971 is still acceptable.Conclusion and measures / preventive measures: based upon the investigation results a clear root cause conclusion cannot be drawn.There is no indication for a material-, manufacturing- or design-related failure.Based upon the investigations results a capa is not necessary.
|
| |
|