(b)(4).Date sent: 5/24/2022.D4: batch #u5ed5y.Investigation summary: the product was returned to ethicon endo-surgery for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the 6r45b (b) reload was received with no apparent damaged.The reload was received fully fired.No functional test could be performed due the condition of the device.Although no conclusion could be reached on the cause of the reported event, the instructions for use do contain the following caution: close the jaws of the instrument by squeezing the closing trigger until it locks in place.An audible click indicates that the closing trigger is locked.Once in the cavity, press the anvil release button to reopen the instrument jaws and return the closing trigger to its original position.To articulate the instrument, turn the articulating lever until it comes to a stop in either direction.To rotate the instrument, turn the rotating knob by applying pressure in a clockwise or counter-clockwise direction.The instrument shaft will rotate freely in either direction.Position the instrument around the tissue to be stapled.The tissue should be positioned between the black line at the distal end of the jaws and the black line at the proximal end, indicating the end of the staple line.The inner black line references the cut line of the device.After positioning the instrument jaws, close the jaws by squeezing the closing trigger until it locks.An audible click indicates that the closing trigger and jaws are locked.Holding the jaws in place for 15 seconds after closing and prior to firing may result in better compression and staple formation.Fire the instrument by squeezing the firing trigger completely with one continuous, smooth stroke until it rests on the closing trigger.To open the jaws, squeeze the closing trigger, and then simultaneously press the anvil release button.While pressure is still on the anvil release button, slowly release the closing trigger.Gently pull the instrument away from the transected tissue and ensure the tissue is released from the jaws.To remove the instrument from the cavity, squeeze the closing trigger until it locks, closing the jaws.Remove the instrument through the trocar in the closed position.The event could not be confirmed as the reload was received fully fired.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.Visual analysis of the returned sample revealed that the 6r45b (c) reload was received with no apparent damaged.The reload was received fully fired.No functional test could be performed due the condition of the device.Although no conclusion could be reached on the cause of the reported event, the instructions for use do contain the following caution: close the jaws of the instrument by squeezing the closing trigger until it locks in place.An audible click indicates that the closing trigger is locked.Once in the cavity, press the anvil release button to reopen the instrument jaws and return the closing trigger to its original position.To articulate the instrument, turn the articulating lever until it comes to a stop in either direction.To rotate the instrument, turn the rotating knob by applying pressure in a clockwise or counter-clockwise direction.The instrument shaft will rotate freely in either direction.Position the instrument around the tissue to be stapled.The tissue should be positioned between the black line at the distal end of the jaws and the black line at the proximal end, indicating the end of the staple line.The inner black line references the cut line of the device.After positioning the instrument jaws, close the jaws by squeezing the closing trigger until it locks.An audible click indicates that the closing trigger and jaws are locked.Holding the jaws in place for 15 seconds after closing and prior to firing may result in better compression and staple formation.Fire the instrument by squeezing the firing trigger completely with one continuous, smooth stroke until it rests on the closing trigger.To open the jaws, squeeze the closing trigger, and then simultaneously press the anvil release button.While pressure is still on the anvil release button, slowly release the closing trigger.Gently pull the instrument away from the transected tissue and ensure the tissue is released from the jaws.To remove the instrument from the cavity, squeeze the closing trigger until it locks, closing the jaws.Remove the instrument through the trocar in the closed position.The event could not be confirmed as the reload was received fully fired.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.Visual analysis of the returned sample revealed that the 6r45b (a) reload was received with no apparent damaged.The reload was received fully fired.No functional test could be performed due the condition of the device.Although no conclusion could be reached on the cause of the reported event, the instructions for use do contain the following caution: close the jaws of the instrument by squeezing the closing trigger until it locks in place.An audible click indicates that the closing trigger is locked.Once in the cavity, press the anvil release button to reopen the instrument jaws and return the closing trigger to its original position.To articulate the instrument, turn the articulating lever until it comes to a stop in either direction.To rotate the instrument, turn the rotating knob by applying pressure in a clockwise or counter-clockwise direction.The instrument shaft will rotate freely in either direction.Position the instrument around the tissue to be stapled.The tissue should be positioned between the black line at the distal end of the jaws and the black line at the proximal end, indicating the end of the staple line.The inner black line references the cut line of the device.After positioning the instrument jaws, close the jaws by squeezing the closing trigger until it locks.An audible click indicates that the closing trigger and jaws are locked.Holding the jaws in place for 15 seconds after closing and prior to firing may result in better compression and staple formation.Fire the instrument by squeezing the firing trigger completely with one continuous, smooth stroke until it rests on the closing trigger.To open the jaws, squeeze the closing trigger, and then simultaneously press the anvil release button.While pressure is still on the anvil release button, slowly release the closing trigger.Gently pull the instrument away from the transected tissue and ensure the tissue is released from the jaws.To remove the instrument from the cavity, squeeze the closing trigger until it locks, closing the jaws.Remove the instrument through the trocar in the closed position.The event could not be confirmed as the reload was received fully fired.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.Visual analysis of the returned sample revealed that the 6r45b (e) reload was received with no apparent damaged.The reload was received fully fired.No functional test could be performed due the condition of the device.Although no conclusion could be reached on the cause of the reported event, the instructions for use do contain the following caution: close the jaws of the instrument by squeezing the closing trigger until it locks in place.An audible click indicates that the closing trigger is locked.Once in the cavity, press the anvil release button to reopen the instrument jaws and return the closing trigger to its original position.To articulate the instrument, turn the articulating lever until it comes to a stop in either direction.To rotate the instrument, turn the rotating knob by applying pressure in a clockwise or counter-clockwise direction.The instrument shaft will rotate freely in either direction.Position the instrument around the tissue to be stapled.The tissue should be positioned between the black line at the distal end of the jaws and the black line at the proximal end, indicating the end of the staple line.The inner black line references the cut line of the device.After positioning the instrument jaws, close the jaws by squeezing the closing trigger until it locks.An audible click indicates that the closing trigger and jaws are locked.Holding the jaws in place for 15 seconds after closing and prior to firing may result in better compression and staple formation.Fire the instrument by squeezing the firing trigger completely with one continuous, smooth stroke until it rests on the closing trigger.To open the jaws, squeeze the closing trigger, and then simultaneously press the anvil release button.While pressure is still on the anvil release button, slowly release the closing trigger.Gently pull the instrument away from the transected tissue and ensure the tissue is released from the jaws.To remove the instrument from the cavity, squeeze the closing trigger until it locks, closing the jaws.Remove the instrument through the trocar in the closed position.The event could not be confirmed as the reload was received fully fired.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.Visual analysis of the returned sample revealed that the 6r45b (d) reload was received with no apparent damaged.The reload was received fully fired.No functional test could be performed due the condition of the device.Although no conclusion could be reached on the cause of the reported event, the instructions for use do contain the following caution: close the jaws of the instrument by squeezing the closing trigger until it locks in place.An audible click indicates that the closing trigger is locked.Once in the cavity, press the anvil release button to reopen the instrument jaws and return the closing trigger to its original position.To articulate the instrument, turn the articulating lever until it comes to a stop in either direction.To rotate the instrument, turn the rotating knob by applying pressure in a clockwise or counter-clockwise direction.The instrument shaft will rotate freely in either direction.Position the instrument around the tissue to be stapled.The tissue should be positioned between the black line at the distal end of the jaws and the black line at the proximal end, indicating the end of the staple line.The inner black line references the cut line of the device.After positioning the instrument jaws, close the jaws by squeezing the closing trigger until it locks.An audible click indicates that the closing trigger and jaws are locked.Holding the jaws in place for 15 seconds after closing and prior to firing may result in better compression and staple formation.Fire the instrument by squeezing the firing trigger completely with one continuous, smooth stroke until it rests on the closing trigger.To open the jaws, squeeze the closing trigger, and then simultaneously press the anvil release button.While pressure is still on the anvil release button, slowly release the closing trigger.Gently pull the instrument away from the transected tissue and ensure the tissue is released from the jaws.To remove the instrument from the cavity, squeeze the closing trigger until it locks, closing the jaws.Remove the instrument through the trocar in the closed position.The event could not be confirmed as the reload was received fully fired.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.Visual analysis of the returned sample determined that the 6r45b reload (f) was received with no apparent damaged and partially fired 1/10 and with the reload lock out spring normal.No functional test could be performed due to the condition of the reload.The condition of the reload indicates that the device's firing cycle was interrupted.When firing the device make sure that the firing stroke is completed.Do not partially fire the device.Fire the device by squeezing the firing trigger completely until it rests on the closing trigger.Once the firing cycle has been initiated, it must be completed.If re-initiation of firing is resumed, the device will lockout.Firing through the lockout mechanism will break the device.Please reference the instruction for use for more information.Visual analysis of the returned sample determined that the 6r45b reload (g) was received with no apparent damaged and partially fired 1/10 and with the reload lock out spring normal.No functional test could be performed due to the condition of the reload.The condition of the reload indicates that the device's firing cycle was interrupted.When firing the device make sure that the firing stroke is completed.Do not partially fire the device.Fire the device by squeezing the firing trigger completely until it rests on the closing trigger.Once the firing cycle has been initiated, it must be completed.If re-initiation of firing is resumed, the device will lock out.Firing through the lockout mechanism will break the device.Please reference the instruction for use for more information.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
|