Model Number 0998-00-3023-53 |
Device Problem
Erratic or Intermittent Display (1182)
|
Patient Problems
Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/03/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
A getinge field service engineer (fse) was dispatched to evaluate this unit.The fse tested the unit and the failure was observed.The fse replaced the display to video cable which was causing the display.The unit passed all functional and safety tests to factory specifications.The unit was cleared for clinical use.A supplemental report will be submitted upon completion of our investigation.
|
|
Event Description
|
It was reported that during the user's staff training the cs300 intra-aortic balloon pump (iabp) had screen blanks and would lash in and out.There was no patient involvement, and no adverse event was reported.
|
|
Manufacturer Narrative
|
A getinge field service engineer (fse) was dispatched to evaluate this unit.Unit was tested and the failure was observed.The fse replaced the display to video cable which was causing the display.The unit passed all functional and safety tests to factory specifications.The unit was cleared for clinical use.
|
|
Search Alerts/Recalls
|