• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CHRISTOPH MIETHKE GMBH & CO KG PROGAV 2.0 WITH SHUNTASSISTANT 30; HYDROCEPHALUS MANAGEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CHRISTOPH MIETHKE GMBH & CO KG PROGAV 2.0 WITH SHUNTASSISTANT 30; HYDROCEPHALUS MANAGEMENT Back to Search Results
Model Number FX415T
Medical Device Problem Code Infusion or Flow Problem (2964)
Health Effect - Clinical Code Hydrocephalus (3272)
Date of Event 01/25/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a progav 2.0 shuntsystem (part # fx415t) was implanted during a procedure performed on (b)(6) 2018.According to the complainant, the shunt system was believed to be operated in overdrainage.The patient underwent a revision procedure on (b)(6) 2022.The complaint device was returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Age: (b)(6).Height: 151 centimeters (cm) weight: (b)(6).Gender: male.
 
Additional Manufacturer Narrative
Investigation: visual inspection: no significant deformations or damage of the valves were detected during the visual inspection.Small particles were visible in the delivery fluid.Permeability test: a permeability test has shown that both valves are permeable.Computer controlled test: to investigate the claim of over-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valves are tested in both the horizontal as well as the vertical positions.The results show that the progav 2.0 is operating within the accepted tolerance in the horizontal position.The shuntassistant is not operating within the specified tolerances in the vertical position.An accelerated outflow could be determined.Adjustment test: the progav 2.0 was tested and is adjustable to all specified pressures.Klick force and brake function test.The brake functionality test has shown that the brake function is fully operational and the braking force is within the given tolerances.Internal inspection : after dismantling of the valves, some deposits were found in both valves.Results: it should be mentioned that after the revision, the permeability test has already been performed by the sender.Based on our investigation, we confirm that the shows an increased flow of liquid at the time of our investigation.This is likely due to the deposits observed inside the valves.As described in scientific literature, the problem encountered is one of the known, inevitable risks of hc-therapy by shunt implants.Deposits caused by natural substances in the cerebrospinal fluid, such as protein, blood or tissue particles, are among the known and unavoidable risks and side effects of hydrocephalus therapy.Small amounts of proteins might compromise the integrity of the valve.We can exclude a defect at the time of release.The shunt system met all specifications of the final inspection when released from christoph miethke (b)(4).No further regulatory actions are required from our point of view.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGAV 2.0 WITH SHUNTASSISTANT 30
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key13662568
Report Number3004721439-2022-00023
Device Sequence Number14704746
Product Code JXG
UDI-Device Identifier04041906136574
UDI-Public4041906136574
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K161853
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/21/2023
Device Model NumberFX415T
Device Catalogue NumberFX415T
Device Lot Number20038292
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/07/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/07/2022
Initial Report FDA Received Date03/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/26/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age13 YR
Patient SexMale
Patient Weight39 KG
-
-