It was reported by the sales rep that during a knee arthroscopy procedure on (b)(6) 2021, it was observed that the distal lens on the arthsco,4/140_30_qik/strkr hub device was in contact with arthroscopic shaver.During in-house engineering evaluation, it was determined that window and optics were cloudy/foggy.Another like device was used to complete the procedure.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.
|
|
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).Investigation summary: the complaint device was received at the service center and evaluated.Per service reports, this complaint can be confirmed.During the service evaluation the following defects were identified: damage to distal tip (bent and burnt), objective, window and optics (cloudy/foggy).Bent/dented - damaged optics.Minor scratches on the unit.The defective parts were replaced to resolve the issues.After repair, the device was found to be working according to the specifications.The faulty parts was identified as the root cause for the device failure during the service evaluation.Manufacturing record evaluation is not required as the reported event is not associated with the manufacturing process and/or the potential cause of the defect cannot be associated to manufacturing.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
|