| Model Number |
68LP-400 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side deflation.Device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual analysis of the returned device identified fold creases, one curved opening and wear abrasion.A microscopic analysis and leak test were performed which identified one opening crease sharp, void >1 mm in non-textured, valve-to shell-bond area and one opening striated.Fill inspection was performed and identified no blockage in the valve.Based on the device analysis the final assessment is: one opening crease sharp in the side posterior assessed as fold flaw opening.One opening striated in the side radius assessed as surgical damage.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side deflation.Device has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: b5, d6(b), h6.
|
| |
|
Event or Problem Description
|
|
Device has been explanted and replaced.
|
| |
|