| Model Number |
C100020 |
| Medical Device Problem Code |
Excess Flow or Over-Infusion (1311)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The sample is reported to be available but has not yet been received by the manufacturer.A review of the device history record is in-progress.All information reasonably known as of 03 mar 2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
|
| |
|
Event or Problem Description
|
|
Fill volume: unknown.Flow rate: unknown.Procedure: unknown.Cathplace: unknown.Infusion start time: unknown.Infusion stop time: unknown.It was reported that the pump finished 10 hours earlier than expected.The pump contained an anti-cancer drug.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: b5 per review of the original information it was determined that two events occurred.The device history record for lot 20066797 was reviewed and the product was produced according to product specifications.All information reasonably known as of 16 mar 2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
|
| |
|
Event or Problem Description
|
|
Avanos medical inc.Received a single report that referenced two different incidences, which were associated with separate units, involving the same patient.This is the first of two reports.Refer to 2026095-2022-00016 for the second report.Per additional information received 4 mar 2022: fill volume: 95ml; flow rate: 2ml/h; specific drug used: 5-fu.
|
| |
|
Additional Manufacturer Narrative
|
|
One used but empty pump was provided for evaluation.When filled with a saline solution, a leak was noted at the filter proximal vent hole.However, flow testing results demonstrated that the flow rate was within specification.No fast flow was observed during testing and the pump operated as expected.The complaint could not be confirmed as reported.No root cause could be determined.All information reasonably known as of (b)(6) 2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
|
| |
|