|
WILLIAM COOK EUROPE COOK CELECT NAVALIGN JUGULAR & FEMORAL VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
|
Back to Search Results |
|
| Model Number |
G52915 |
| Medical Device Problem Codes |
Structural Problem (2506); Device Tipped Over (2589); Unintended Movement (3026)
|
| Health Effect - Clinical Codes |
Internal Organ Perforation (1987); Perforation of Vessels (2135); Anxiety (2328); Ambulation or Postural Difficulties (2544)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Occupation: non-healthcare professional.Investigation: the reported allegations have been further investigated based on the information provided to date.The following allegations have been investigated: organ/ vena cava perforation.Filter interacts with ivc wall, e.G.Penetration/perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma, vena cava perforation, vena cava penetration.Catalog number and lot number are unknown.The alleged celect is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information becomes available.
|
| |
|
Event or Problem Description
|
|
The following information is alleged: the patient received a celect inferior vena cava (ivc) filter on (b)(6) 2012.The patient experienced multiple perforations of the ivc by the struts of the filter.Struts are further perforating and abutting other organs in patient's body.Hospital and medical records have been requested, but not yet provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation: investigation is reopened due to additional information provided.The following allegations have been investigated: organ/ vena cava perforation, anxiety, worry, fear, limited physical activity.The reported allegations have been further investigated based on the information provided to date.Filter interacts with ivc wall, e.G.Penetration/perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma, vena cava perforation, vena cava penetration.Unknown if the reported anxiety, worry, fear, and limited physical activity are directly related to the filter and unable to identify a corresponding failure mode at this point in time.(b)(4) devices in lot.To date, two other unrelated complaints have been reported against the lot.The associated work order was reviewed.No related/relevant notes were documented.The device is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
| |
|
Event or Problem Description
|
|
Patient allegedly received an implant on (b)(6) 2012 via the right common femoral vein due to bariatric surgery.Patient is alleging vena cava and organ perforation, perforation that abuts the transverse duodenum, and perforation that abuts the posterior aorta.Patient further alleges "i live with the anxiety of having a filter that could fail further at any time, but that cannot be retrieved without a serious surgery.I am worried because my filter has perforated outside of my vena cava and is abutting adjacent structures." "for fear of aggravating my current condition and causing potential future injuries related to the cook filter, i try not to engage in any activities that could cause me to exert myself.This includes standing and walking for long periods of time.".
|
| |
|
Event or Problem Description
|
|
Per computed tomography report, "caval perforation: yes.Grade 4 perforation of the 6 o'clock strut into the right psoas muscle.Grade 3 perforation of the 10, 11, 12 and 1 o'clock struts to abut transverse duodenum.Grade 3 perforation of 4 o'clock strut to abut posterior aorta.Tilt: yes.7.15 degrees of coronal and 10.46 degrees of sagittal tilt.Apex of filter touches right anterior wall of ivc at level of right renal vein confluence.Migration: yes.Apex of filter at level of right renal vein confluence.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation: investigation is reopened due to additional information provided.The following allegations have been investigated: migration, tilt.The reported allegations have been further investigated based on the information provided to date.Filter or filter fragment migration and (or) embolization (e.G., movement to the heart or lungs) has been reported.Filter or filter fragment movement has occurred in both the cranial and caudal direction and may be either symptomatic or asymptomatic.Potential causes may include filter placement in ivcs with diameters smaller or larger than those specified in these instructions for use; improper deployment; deployment into thrombus; dislodgement due to large thrombus burdens; and (or) excessive force or manipulations near an in situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: filter migration, trauma to adjacent structures.Filter tilt has been reported.Potential causes may include filter placement in ivcs with diameters larger than those specified in these instructions for use; improper deployment; manipulations near an implanted filter (e.G., a surgical or endovascular procedure in the vicinity of a filter); and (or) a failed retrieval attempt.20 devices in lot.Two other complaints have been reported against the lot for different issue.The associated work order was reviewed.No related/relevant notes were documented.The device is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
| |
|
Search Alerts/Recalls
|
|
|