| Catalog Number |
07.02100.001 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/31/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Inspection: the returned device was examined.Visual inspection revealed that the center screw was fractured and disassembled from the device.Dhr review: the dhr was reviewed.There are no indications of manufacturing issues which would have contributed to this event and the device was likely conforming when it left zimvie¿s control.Potential root cause: a definitive root cause cannot be determined with the information provided.This event could possibly be attributed to wear through use over time or excessive force applied to the instrument.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported that the center screw of a vitality rod gripper broke while surgeon was using it on a rod during posterior spinal fusion.There was no reported patient harm and all pieces were accounted for.
|
| |
|
Search Alerts/Recalls
|