According to the reporter, during thoracoscopic partial pulmonary resection, when resecting the pulmonary parenchyma, an abnormal noise was heard from the reload and adapter.Upon checking, the part of the reload connecting to the adapter broke off.A competitor device was used to resolve the issue in order to complete the case.There was no patient injury.
|
|
Additional information: g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the reload was fully fired with the interlock engaged, the jaw of the reload was open, and the adapter was broken.Functionally, the adapter was bent but the reload was loaded into a representative handle.The reload could not articulate due to breakage of the adapter.It was reported that the device broke.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: ensure that the black return knob on the instrument is pulled back completely and the articulation lever is neutral (0° relative) to the instrument.Cycle the instrument to confirm proper loading of the reload.To do so, squeeze the handle once to close the jaws of the reload.Push the black loop handle all of the way forward or pull back on the black return knob and confirm that the reload jaws open fully.Failure to completely fire the reload will result in an incomplete cut and/or incomplete staple formation, which may result in poor hemostasis and/or leakage.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|