• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION SYNERGY XD; CORONARY DRUG-ELUTING STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION SYNERGY XD; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number H7493941816400
Medical Device Problem Codes Failure to Advance (2524); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/14/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that stent damage occurred.The target lesion was located in a coronary artery.A 4.00 x 16mm synergy xd drug-eluting stent was advanced for treatment.However, the device failed to cross the lesion and the proximal struts of the stent were bent.The procedure was completed with another of the same device.There were no patient complications reported and the patient status was stable.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: a synergy xd mr us 4.00 x 16mm stent delivery system was returned for analysis.A visual examination of the stent found distal stent damage, with distal stent struts lifted.The undamaged stent outer diameter was measured within maximum crimped stent profile measurement.The balloon cones were reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip found no issues.A visual and tactile examination of the hypotube found multiple hypotube kinks at several locations along the hypotube shaft.A visual examination of the outer and inner lumen and mid-shaft section found no issues along the shaft polymer extrusion.No other issues were identified during the product analysis.
 
Event or Problem Description
It was reported that stent damage occurred.The target lesion was located in a coronary artery.A 4.00 x 16mm synergy xd drug-eluting stent was advanced for treatment.However, the device failed to cross the lesion and the proximal struts of the stent were bent.The procedure was completed with another of the same device.There were no patient complications reported and the patient status was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY XD
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key13664854
Report Number2134265-2022-02425
Device Sequence Number4755707
Product Code NIQ
UDI-Device Identifier08714729981121
UDI-Public08714729981121
Combination Product (Y/N)Y
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/13/2023
Device Model NumberH7493941816400
Device Lot Number0028129499
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/22/2022
Supplement Date Received by Manufacturer04/13/2022
Initial Report FDA Received Date03/03/2022
Supplement Report FDA Received Date04/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/13/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-