| Model Number |
45-181000 |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/03/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a partial splenic embolization coil, an azur vascular plug was implanted.Eight minutes after implantation, the plug was observed to have moved 1-2 cm from the original position.The case with successfully completed.
|
| |
|
Additional Manufacturer Narrative
|
|
A search for non-conformances associated with this part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.Although the device was implanted in the patient, images were provided for analysis.The investigation is currently ongoing.
|
| |
|
Additional Manufacturer Narrative
|
|
No product(s) were returned for evaluation, but two images were provided.The first image shows the two radiopaque markers of the azur vascular plug in the mid-portion of the splenic artery.The second image shows that the plug markers have migrated distally approximately 2 cm.These images confirm the complaint and show the plug migrated distal to the original deployment.Without the return and physical evaluation of the device, the investigation is unable to determine if a condition existed that would have caused or contributed to the reported event.
|
| |
|