| Model Number |
212480 |
| Medical Device Problem Code |
Residue After Decontamination (2325)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/08/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Angled and straight saw attachments leaking oil.No surgical delay.Case type / application: tka.
|
| |
|
Additional Manufacturer Narrative
|
|
As part of normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplemental report will be submitted when additional information becomes available.Not available.
|
| |
|
Event or Problem Description
|
|
Angled and straight saw attachments leaking oil.No surgical delay.Case type / application: tka.
|
| |
|
Additional Manufacturer Narrative
|
|
Update to lot/serial no.& mfg date.Reported event: an event regarding foreign matter involving a mako saw attachment was reported.The event was not confirmed because the product was not available for inspection.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: the alleged failure mode was not confirmed because the product was not available for inspection.No additional investigation or specific actions are required.If additional information is received then the complaint will be reopened.
|
| |
|
Search Alerts/Recalls
|