• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC 3MM FLUTED MATCHSTICK, 15CM; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC 3MM FLUTED MATCHSTICK, 15CM; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) Back to Search Results
Model Number MA15-8NS
Medical Device Problem Codes Material Fragmentation (1261); Separation Problem (4043)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/08/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during testing, it was discovered that the attachment device had ¿sticky burrs¿.It was further reported that the attachment came back with a cutter fragment stuck in it and nothing else was visible.The event did not occur during surgery.There was no patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Additional Manufacturer Narrative
Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.Concomitant med products and therapy dates: attachment device, (b)(6) 2022.The serial number was unknown; therefore, the device manufacture date is unknown, and the udi is incomplete.The actual device has been returned and is currently pending evaluation.Once the investigation has been completed, and if additional information should become available, a supplemental medwatch report will be submitted accordingly.Udi: (b)(4).
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.Device evaluation: the actual device was returned for evaluation.Quality engineering visually inspected the drill bit device and found that it failed visual inspection because the device¿s tip had broken off and a piece of the drill bit device was found.The attachment device was taken apart and it was noted that it had excessive corrosion, and medical debris in the locking mechanism assembly, preventing the lock tabs from releasing the cutters which caused it to become stuck.The piece of the cutter was examined using 25x magnification and rechecked on an optical comparator.A full assessment of the device could not be performed due to the condition the cutter was received.The piece of the cutter had broken off below the laser marking and identification of the cutter and the lot number was not able to be identified and the rest of the cutter was not sent in.It was further determined that the cutter getting stuck was likely due to debris and residue preventing the lock tabs from moving into place.The cause of the medical debris and detergent residue buildup was due to improper cleaning (user error).Therefore, the reported condition of a cutter fragment stuck in the attachment was confirmed.The assignable root cause was determined to be traced to maintenance, which is improper maintenance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3MM FLUTED MATCHSTICK, 15CM
Common Device Name
DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
kara ditty-bovard
4500 riverside drive
palm beach gardens, FL 33410
6103142063
MDR Report Key13666283
Report Number1045834-2022-00274
Device Sequence Number4756313
Product Code HBE
UDI-Device Identifier00845384004615
UDI-Public00845384004615
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K042783
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMA15-8NS
Device Catalogue NumberMA15-8NS
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/17/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/28/2022
Supplement Date Received by Manufacturer06/07/2022
Initial Report FDA Received Date03/04/2022
Supplement Report FDA Received Date06/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-