| Model Number |
BRX500S18 |
| Medical Device Problem Code |
Degraded (1153)
|
| Health Effect - Clinical Code |
Nodule (4551)
|
| Date of Event |
02/22/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The manufacturer received information alleging an issue related to a continuous positive airway pressure (cpap) device's sound abatement foam became degraded and caused the patient to develop lung nodule.The medical intervention that the patient received in response to the event is currently unknown.This issue was reported to the fda per 21 cfr 806.The device will be corrected per res 88058.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturer was previously contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer received information alleging the patient had developed lung nodule.The medical intervention is unknown.Despite the three attempts to have the device returned for evaluation and investigation were unsuccessful.The device has not yet been returned to the manufacturer for evaluation.At this time, no further investigation can be performed.If any additional information is received, a follow up report will be filed.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer previously received information alleging the patient had developed lung nodule related to a cpap device's sound abatement foam.The medical intervention is unknown.The device was returned to the manufacturer's service center for further evaluation.The device was evaluated.There was no mention of visual findings to the external part of the device.The internal aspect of the device was inspected.The device powered on and airflow was confirmed.The device's downloaded logs were reviewed by the manufacturer.There were no errors found.The manufacturer concludes that they could not confirm the customer's allegation and there was no visible foam degradation and unit scrapped.In this report, section d8, d9, h2, h3 and h6 has been updated.
|
| |
|
Search Alerts/Recalls
|