It was reported that, during a knee arthroscopy, the biorci-ha implant broke off for no apparent reason.The implant did not break inside the patient and there was no void left in the patient.Surgery was resumed, after a non-significant delay, with a back-up device.No patient complications were reported.
|
|
H3, h6: the reported device was received for evaluation.It was determined the device did not contribute to the reported event.A visual inspection of the returned device found that it is not in its original packaging.The screw is broken on the distal end, and there is debris on the device.Based on the condition of the product material found during visual inspection, additional material testing is not required.The complaint was confirmed, and the root cause was associated with unintended use of the device.Factors that could have contributed to the reported event include excessive force on the device, excessive torque on the device, attempted correction of a damaged device, off-axis insertion, improper preparation of the insertion site, or an inadvertent impact event inconsistent with normal use.A complaint history review found similar reported events.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the polymer found that the storage requirements for the material are specified, and the material must comply with provided percentage composition specifications as measured by ash test.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.
|