Model Number 072189
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/03/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that the edge of the silicone drain was broken off.
 
Additional Manufacturer Narrative
The reported event was confirmed cause unknown.1 sample were confirmed to exhibit the reported failure.The device had not met specifications.The product was not used for patient treatment.The product caused the reported failure.Visual evaluation of the returned sample noted one opened (without original packaging), silicone channel drain with trocar.Visual inspection of the sample noted the trocar point shield the broken off upon return.This does not meet specification per ip7600518, revision 21 which states "no cuts or tears are allowed on the surface of the drains".A potential root cause for this failure mode could be ¿drain was manufactured incorrectly (dimensions or material out spec, curing time not appropriate".The device history record review could not be performed without a lot number.The investigation is concluded, and no additional action is required at this time.The device was returned for evaluation.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "indications: bard® channel drains, round and flat silicone, are indicated for use with selected bard evacuators for closed wound drainage following head and neck, orthopedic, abdominal, ent, ob/gyn, plastic, neurosurgery, thoracic and cardiovascular (channel drains only) procedures.Warning: when placing drain(s) care should be taken to ensure that the channeled portion of the wound drain lies completely within the confines of the wound.Read product insert provided with the closed wound evacuator for detailed instructions, warnings and precautions associated with the use of the evacuator device.To avoid the possibility of drain damage or breakage: avoid suturing through drains.Drains should lie flat and in line with the skin exit areas.Particular care should be taken to avoid any obstacles to the drain exit path.Drains should be checked for free motion during closure to minimize the possibility of breakage.Drain removal should be done gently by hand.Drains should not be handled with pointed, toothed or sharp instruments which could cause cuts or nicks and lead to subsequent structural failure of the drain.Surgical removal may be necessary if drain is difficult to remove or breaks.Drain placement a.Using a single silicone channel drain 1.Place channeled wound drain within critical fluid collection area of wound.2.For drains displaying numerical depth markers, draw non-channeled section of wound drain through to the outside until desired depth has been achieved.For drains without numerical depth markers, draw non-channeled section of wound drain through to the outside until desired drain depth has been achieved or until drain indicator dot appears at the skin surface.Drains displaying numerical depth markers are provided with an additional radiopaque ring to indicate the transition into the channeled portion.These radiopaque indications will appear perpendicular to each other under x-ray.(see ¿warning¿ above).3.Insert connecting tube of drain to a bard brand evacuator.Included adaptors may be required for attachment.B.Using two silicone channel drains 1.Repeat steps 1-2 above.2.Attach proximal ends of silicone drains to a bard y-connector and connect y-connector to evacuator drain port.Included adaptors may be required for attachment." h11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.H3 other text: the actual/suspected device was inspected.
 
Event or Problem Description
It was reported that the edge of the silicone drain was broken off.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SILICONE CHANNEL DRAIN WITH 3/16 IN. TROCAR, 15 FR., ROUND, FULL FLUTED
Common Device Name
SILICONE CHANNEL DRAIN
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
juan velez
8195 industrial blvd
covington 30014
7707846100
MDR Report Key13666778
Report Number1018233-2022-01046
Device Sequence Number5466454
Product Code GBX
UDI-Device Identifier00801741049712
UDI-Public(01)00801741049712
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number072189
Device Catalogue Number072189
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/28/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/07/2022
Supplement Date Received by Manufacturer08/19/2022
Initial Report FDA Received Date03/04/2022
Supplement Report FDA Received Date08/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-