MAUDE Adverse Event Report: ZIMMER SWITZERLAND MANUFACTURING GMBH CORTICAL BONE SCREW, 4X26MM; AFFIXUS NATURAL NAIL SYSTEM HUMERAL NAIL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SWITZERLAND MANUFACTURING GMBH CORTICAL BONE SCREW, 4X26MM; AFFIXUS NATURAL NAIL SYSTEM HUMERAL NAIL Back to Search Results
Model Number N/A
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 02/07/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that operation was performed with ann nail system.After 4 months from the initial, surgeon found a screw of the proximal screws was backed out from the proper position.The surgeon keep an eye on the patient condition as well as no revision will be planned so far.Patient involvement.No further outcome.
 
Additional Manufacturer Narrative
Concomitant medical products: blunt tip screw, 4x44mm; catalog#: 47-2486-044-40; lot#: 3010653.Blunt tip screw, 4x46mm; catalog#: 47-2486-046-40; lot#: 3024714.Blunt tip screw, 4x60mm; catalog#: 47-2486-060-40; lot#: 3054546.Proximal humerus nail cap, 0mm; catalog#: 47-2488-010-00; lot#: 3033538.The manufacturer received x-rays and will be reviewed as part of ongoing investigation.The manufacturer did not receive the device for investigation.Where lot numbers were received for the devices, the device history records were reviewed and found to be conforming.A cause for this specific event cannot be ascertained from the information provided.As soon as supplemental information becomes available an updated report will be submitted.Zimmer biomet¿s reference number of this file is (b)(4).The following reports are associated with this event: 0009613350-2022-00126; 0009613350-2022-00128; 0009613350-2022-00129.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay corrected information investigation results are now available.During the investigation it was concluded that the device previously submitted on this medwatch is not a contributing element for the reported event and therefore has been moved to concomitant device.Please remove 0009613350-2022- 00127 from your records.Multiple mdr reports were filed for this event, please see associated reports: 0009613350-2022-00126-1; 3068920 0009613350-2022-00129-1; 0009613350-2022- 00216; 0009613350-2022-00217.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
Investigation results are now available.During the investigation it was concluded that the device previously submitted on this medwatch is not a contributing element for the reported event and therefore has been moved to concomitant device.Please remove 0009613350-2022- 00127 from your records.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CORTICAL BONE SCREW, 4X26MM
Common Device Name
AFFIXUS NATURAL NAIL SYSTEM HUMERAL NAIL
Manufacturer (Section D)
ZIMMER SWITZERLAND MANUFACTURING GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ  8404
Manufacturer (Section G)
ZIMMER SWITZERLAND MANUFACTURING GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ   8404
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key13666847
Report Number0009613350-2022-00127
Device Sequence Number14641638
Product Code HSB
UDI-Device Identifier00889024505339
UDI-Public00889024505339
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K200814
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2026
Device Model NumberN/A
Device Catalogue Number47-2486-126-40
Device Lot Number3068970
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/11/2022
Supplement Date Received by Manufacturer04/01/2022
Initial Report FDA Received Date03/04/2022
Supplement Report FDA Received Date04/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/09/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10 NARRATIVE
Outcome Attributed to Adverse Event Hospitalization;
Patient SexFemale
-
-