Model Number FV441T
Medical Device Problem Code Infusion or Flow Problem (2964)
Health Effect - Clinical Code Hydrocephalus (3272)
Date of Event 02/09/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Visual inspection: the following observations were made during the visual inspection: deposits on the catheter.Permeability test: the test showed that the progav shunt system is permeable.The liquid was bloody.Computer control test: the computer controlled test showed the opening pressure of the progav, at a reference flow rate of 20 ml/h in a horizontal position of the valve, to be 17,07 cmh2o.This is within the specified tolerance of 14 cmh2o ± 3 cmh2o.An applied pressure of 14 cmh2o, with the device in the horizontal position is expected to have a resultant pressure of 14 cmh2o ± 3 cmh2o.Additionally, the shuntassistant was tested according to standard procedure in the vertical position of the valve.At a fixed opening pressure of 20 cmh2o in the vertical position, a pressure of 20 cmh2o -2/+4 cmh2o is expected.The results indicated that at a reference flow of 20 ml/h in the vertical position, the shuntassistant had a pressure of 23,33 cmh2o.This is within the specified tolerance of 20 cmh2o -2/+4 cmh2o.Adjustability test: the progav was found to be adjustable to all pressure settings.The shuntassistant is a fixed pressure valve, therefore the adjustability test is inapplicable.Braking force and brake function test: the braking force of the progav was within the specified tolerance and the brake function operated as expected.Internal inspection of product: after dismantling of the valves, deposits were found in both valves.Results: based on our investigation results, we cannot identify a malfunction in the shunt system.However, we assume that the significant deposits found within the valve may have temporarily caused the functional impairment.Deposits caused by natural substances in the cerebrospinal fluid, such as protein, blood or tissue particles, are among the known and unavoidable risks and side effects of hydrocephalus therapy.Small amounts of non-visible deposits/proteins might compromise the integrity of the valve.We can exclude a defect at the time of release.The shunt system met all specifications of the final inspection when released from christoph miethke (b)(4).From our point of view, no further regulatory actions are required.
 
Event or Problem Description
It was reported that a progav shunt system (part # fv441t) was implanted during a procedure performed on (b)(6) 2009.According to the complainant, the shunt system was believed to be clogged.The patient underwent a revision procedure on (b)(6) 2022.The complaint device was returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Age: (b)(6).Height: 120 centimeters (cm).Weight: (b)(6).Gender: male.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGAV SYS PED.W/SA 20 A.PRECHAMBER
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key13666852
Report Number3004721439-2022-00050
Device Sequence Number4756332
Product Code JXG
UDI-Device Identifier04041906132545
UDI-Public4041906132545
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K103003
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFV441T
Device Catalogue NumberFV441T
Device Lot Number20007802
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/22/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/10/2022
Initial Report FDA Received Date03/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/11/2008
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age12 YR
Patient SexMale
Patient Weight21 KG
-
-