| Model Number |
ASD37A |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/17/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported the physician selected a 37mm gore® cardioform asd occluder to treat an atrial septal defect with a static measurement of 23mm x 24mm (balloon sizing was not performed) and a septal length of 34mm x 47mm.The occluder was deployed and locked with good septal apposition.A minor residual shunt was noted.The following day, imaging showed the occluder had embolized to the left pulmonary artery.This device was snared and a 44mm gore® cardioform asd occluder was implanted with no adverse effects.The patient was doing well following the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
The gore® cardioform asd occluder instructions for use list device embolization as a potential clinical and device adverse event.
|
| |
|