| Model Number |
IPN043837 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
02/10/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that "torn product found while connecting breathing circuit to air way".No additional information provided at time of report.Patient condition unknown at time of report.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that "torn product found while connecting breathing curcuit to air way".No additional information provided at time of report.Patient condition unknown at time of report.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The sample was returned for evaluation.A visual exam was performed and it was observed that the corrugated tubing was torn at the connector area.A device history record review was performed on a potential lot number reported by the customer (lot #kmh21j0110) and no relevant findings were identified.Based on the visual exam, the reported complaint was confirmed.The manufacturing site reports "the defect part was supplied by our supplier." a non-conformance was opened to address this issue.
|
| |
|