• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AVANOS MEDICAL INC. HOMEPUMP C-SERIES, 270 ML, 5 ML/HR (CONTAINS DEHP); ELASTOMERIC LFR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AVANOS MEDICAL INC. HOMEPUMP C-SERIES, 270 ML, 5 ML/HR (CONTAINS DEHP); ELASTOMERIC LFR Back to Search Results
Model Number C270050
Medical Device Problem Code Excess Flow or Over-Infusion (1311)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 02/09/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product involved in the report has been returned and is being processed for evaluation.A review of the device history record is in-progress.All information reasonably known as of 03-mar-2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).
 
Event or Problem Description
Fill volume: 241ml, flow rate: 5ml/hr, procedure: unknown, cathplace: unknown.It was reported the "patient called pharmacist to let them know that the pump infusion was complete," the pump infused over 23.5 hours instead of the scheduled 72-hours." the patient was admitted into the hospital, but the pharmacist was unaware of what symptoms were experienced.The patient is receiving an antidote while in the hospital.
 
Additional Manufacturer Narrative
The device history record for the reported lot number 30163048, in this complaint was reviewed and the material was produced according to the manufacturing procedures and met the quality requirements.The device was evaluated and the investigation determined the device was within specification.The reported fast flow was not experienced during the flow accuracy testing and pressure pot testing.There was an obvious leakage at the filter and the filter was removed prior to flow accuracy and pressure pot testing.Root cause could not be determined.All information reasonably known as of 15-apr-2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOMEPUMP C-SERIES, 270 ML, 5 ML/HR (CONTAINS DEHP)
Common Device Name
ELASTOMERIC LFR
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
Manufacturer (Section G)
AVENT S. DE R.L. DE C.V.
ave noruega edificio d-1b
fraccionamiento rubio
tijuana b.c.
Manufacturer Contact
lisa clark
5405 windward parkway
alpharetta, GA 30004
4704485444
MDR Report Key13668120
Report Number2026095-2022-00009
Device Sequence Number17366929
Product Code MEB
UDI-Device Identifier10680651135688
UDI-Public10680651135688
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K052117
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date (Section B) 04/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/30/2024
Device Model NumberC270050
Device Catalogue NumberN/A
Device Lot Number30163048
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/21/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/10/2022
Supplement Date Received by Manufacturer03/21/2022
Initial Report FDA Received Date03/04/2022
Supplement Report FDA Received Date04/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/27/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
5-FU/NS
Outcome Attributed to Adverse Event Other;
Patient Age63 YR
Patient SexMale
Patient Weight83 KG
-
-