| Model Number |
777620 |
| Medical Device Problem Code |
Deformation Due to Compressive Stress (2889)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a black thread knot was found in the stent.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that a black thread knot was found in the stent.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event is unconfirmed as the device met product specifications.The ureteral stent and pusher were returned in the original opened packaging.An evaluation was completed by futurematrix on 02may2022.The only knot present in the returned sample was the one instructed to be there which states: "tie loose ends using surgeons knots, making sure that no loops are left visible within the knot." additionally states, "suture loop attached with single loop method." no defects were detected in the suture and no additional knots were present.As the reported event is unconfirmed a dhr review is not required.As the reported event is unconfirmed, a label/packaging review is not required.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
|
| |
|