Model Number TJM4011
Medical Device Problem Code Contamination (1120)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/17/2022
Type of Reportable Event Malfunction
Event or Problem Description
Where during the control a blue plastic debris was noticed.He observed the presence of "blue" colored particles (inside the individual packaging) which would appear to be "debris" of blue plastic from the device.The manufacturer is required to justify the presence and harmlessness of these "debris".
 
Additional Manufacturer Narrative
(b)(4) initial emdr submission.A follow up emdr will be submitted if additional information becomes available.(b)(4).
 
Additional Manufacturer Narrative
(b)(4) follow-up emdr for device evaluation: five photos were provided to our quality team for investigation.Through visual inspection of the photos indicate blue foreign matter that is larger than 0.80 square millimeters, the cut off point for flash size.The size of these particles is over the limit when checked during the tappi particle size evaluation therefore, the reported failure mode was confirmed.A review of the internal manufacturing device records and raw material history files for the reported lot number 0001394942 was performed and no recorded quality problems or rejections related to this incident were found.Product undergoes inspections during manufacturing, no issues related to the reported incident were identified, all procedural and functional requirements for product release have been met.Based on the information provided, a probable root cause is inattention to detail for this investigation.Complaints received for this device and defect will continue to be monitored by our quality team for signs of emerging trends.H3 other text : see manufacturer here.
 
Event or Problem Description
Where during the control a blue plastic debris was noticed.He observed the presence of "blue" colored particles (inside the individual packaging) which would appear to be "debris" of blue plastic from the device.The manufacturer is required to justify the presence and harmlessness of these "debris".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JAMSHIDI (TJ) NEEDLE BM 11G X3.5 ASP MAC
Common Device Name
INSTRUMENT, BIOPSY
Manufacturer (Section D)
CAREFUSION, INC
400 east foster rd
mannford OK 74044
Manufacturer (Section G)
CAREFUSION, INC
400 east foster rd
mannford OK 74044
Manufacturer Contact
anna wehrheim
75 n. fairview drive
vernon hills, IL 60061
8015652341
MDR Report Key13670643
Report Number1625685-2022-00034
Device Sequence Number12011060
Product Code KNW
UDI-Device Identifier10885403044588
UDI-Public(01)10885403044588
Combination Product (Y/N)N
Initial Reporter CountryTS
PMA/510(K) Number
K171531
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2025
Device Model NumberTJM4011
Device Catalogue NumberTJM4011
Device Lot Number0001394942
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/17/2022
Supplement Date Received by Manufacturer04/15/2022
Initial Report FDA Received Date03/04/2022
Supplement Report FDA Received Date04/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-