Model Number MX448HFB
Medical Device Problem Code No Flow (2991)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the iv fluid is not running through the administration set.There was no patient injury reported.
 
Additional Manufacturer Narrative
One filter set was received for evaluation in relation to the reported event.Evaluation of the device is currently in progress.Once the evaluation is complete, a follow-up report with evaluation results will be submitted.
 
Additional Manufacturer Narrative
This mdr was generated under protocol b10010116, as a result of warning letter cms# (b)(4).A product sample was received for evaluation.Visual and functional testing were performed.Sample went through water leak and occlusion test and it did not exhibit leakage or occlusion condition.The root cause of the reported issue cannot be determined, as the returned sample did not exhibited occlusion nor defective condition.Actions were taken to mitigate the reported issue: awareness meeting with operators and quality inspectors about the failure mode reported on this complaint was held.No problems or issues were identified during this device history record review.
 
Additional Manufacturer Narrative
Other text: this mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).A product sample was received for evaluation.Visual and functional testing were performed.Sample went through water leak and occlusion test and it did not exhibit leakage or occlusion condition.The root cause of the reported issue cannot be determined, as the returned sample did not exhibited occlusion nor defective condition.Actions were taken to mitigate the reported issue: awareness meeting with operators and quality inspectors about the failure mode reported on this complaint was held.No problems or issues were identified during this device history record review.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDEX FILTER SET
Common Device Name
FILTER, INFUSION LINE
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6250 shier rings road
dublin OH 43016
Manufacturer (Section G)
NULL
6250 shier rings road
dublin OH 43016
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key13672452
Report Number3012307300-2022-04347
Device Sequence Number5375485
Product Code FPB
UDI-Device Identifier10351688507457
UDI-Public10351688507457
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K860803
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/27/2024
Device Model NumberMX448HFB
Device Catalogue NumberMX448HFB
Device Lot Number4180739
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/22/2022
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/04/2022
Supplement Date Received by Manufacturer02/03/2023
02/03/2023
Initial Report FDA Received Date03/04/2022
Supplement Report FDA Received Date03/15/2023
03/15/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/01/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-