• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOVASCULAR LLC VASOVIEW 7 XB BISECTOR (OUS); ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOVASCULAR LLC VASOVIEW 7 XB BISECTOR (OUS); ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number VASOVIEW 7 XB BISECTOR (OUS)
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/07/2022
Type of Reportable Event Malfunction
Event or Problem Description
The hospital reported that during an endoscopic vein harvesting procedure using vasoview 7 xb bisector (ous), the short port blunt tip trocar (btt) was discovered to be cracked when opened the vv7.They opened another unit of vv7 and finished the surgery.No patient involvement.
 
Additional Manufacturer Narrative
Trackwise id (b)(4).The device has not yet been returned to maquet cardiac surgery for evaluation.We are following up with the customer for the return of the device.A supplemental report will be submitted if the device is received.
 
Additional Manufacturer Narrative
The device has been returned to the factory and is being evaluated.A supplemental report will be submitted when the evaluation is completed.Trackwise id#: (b)(4).
 
Event or Problem Description
The hospital reported that during an endoscopic vein harvesting(ous), the short port blunt tip trocar (btt) was discovered to be cracked when opened the vv7.They opened another unit of vv7 and finished the surgery.No patient involvement.
 
Additional Manufacturer Narrative
Trackwise # (b)(4).The lot # 25158845 history record review was completed.There were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.The device was returned to the factory for evaluation on 06/16/2022.Photographs were provided by the account.A photographic inspection was conducted.Signs of clinical use and evidence of blood was observed.A crack was observed on the inside of the gray endoscope seal.No other visual defects were observed.An investigation was conducted on 07/20/2022.Signs of clinical use and no evidence of blood was observed.There was a crack observed on the inside of the gray endoscope seal.No other visual defects were observed.Based on the photographic inspection as well as the investigation results, the reported failure "crack" was confirmed.For the reported failure "crack" the root cause is undetermined because the event occurred during the use of the device and the procedural operation is unavailable for our review.The certificate of conformance (c of c) was reviewed.The vendor certifies that this device lot conforms to all applicable product specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VASOVIEW 7 XB BISECTOR (OUS)
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer (Section G)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 
MDR Report Key13673689
Report Number2242352-2022-00194
Device Sequence Number5302812
Product Code GEI
UDI-Device Identifier00607567700437
UDI-Public00607567700437
Combination Product (Y/N)N
Initial Reporter CountryHK
PMA/510(K) Number
K041981
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/14/2023
Device Model NumberVASOVIEW 7 XB BISECTOR (OUS)
Device Catalogue NumberVH-3200-W
Device Lot Number25158845
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/07/2022
Supplement Date Received by Manufacturer06/16/2022
08/08/2022
Initial Report FDA Received Date03/04/2022
Supplement Report FDA Received Date07/06/2022
08/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/14/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNKNOWN.
Patient SexUnknown
-
-