• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TORNIER S.A.S. S40 CORTILOC PEGGED GLENOID TRIAL; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TORNIER S.A.S. S40 CORTILOC PEGGED GLENOID TRIAL; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Catalog Number MWE403
Medical Device Problem Codes Break (1069); Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/04/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Once the investigation has been completed any additional information will be reported in a supplemental report.
 
Event or Problem Description
It was reported that the device broke while inserting in the patient body during the procedure.No additional information is available.
 
Additional Manufacturer Narrative
The reported event could be confirmed, since the product was returned for evaluation and matches the alleged failure mode.The device inspection revealed the following: the received device shows breakage in one of the three pegs indicating intensive usage.Grasping area is deformed with grasper marks due to an application of high external forces.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.Based on the investigation, the root cause was attributed to a user related issue.The device breakage is due to an application of high external forces.If any further information is provided, the investigation report will be reassessed.
 
Event or Problem Description
It was reported that the device broke while inserting in the patient body during the procedure.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S40 CORTILOC PEGGED GLENOID TRIAL
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
TORNIER S.A.S.
161 rue lavoisier
montbonnot saint-martin 38330
FR  38330
Manufacturer (Section G)
TORNIER S.A.S.
161 rue lavoisier
montbonnot saint-martin 38330
FR   38330
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key13674733
Report Number3000931034-2022-00140
Device Sequence Number4996061
Product Code LXH
UDI-Device Identifier03700386974410
UDI-Public03700386974410
Combination Product (Y/N)N
Initial Reporter StateIA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberMWE403
Device Lot Number20C389
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/02/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/08/2022
Supplement Date Received by Manufacturer05/09/2022
Initial Report FDA Received Date03/04/2022
Supplement Report FDA Received Date06/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/11/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-