U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W. L. GORE & ASSOCIATES, INC. GORE® VIABAHN® ENDOPROSTHESIS WITH HEPARIN BIOACTIVE SURFACE; STENT, SUPERFICIAL FEMORAL ARTERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W. L. GORE & ASSOCIATES, INC. GORE® VIABAHN® ENDOPROSTHESIS WITH HEPARIN BIOACTIVE SURFACE; STENT, SUPERFICIAL FEMORAL ARTERY Back to Search Results
Model Number VBHR060502A
Medical Device Problem Codes Loss of or Failure to Bond (1068); Material Separation (1562); Detachment of Device or Device Component (2907); Separation Problem (4043)
Health Effect - Clinical Codes Device Embedded In Tissue or Plaque (3165); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/07/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional manufacturer narrative: cbas® heparin surface incorporates cbas-heparin manufactured from heparin sodium api, which is covalently bound to the device surface and is essentially non-eluting.
 
Event or Problem Description
The following information was reported to gore: on (b)(6) 2022, the patient was being treated for a stenotic lesion in a iliac artery using a gore® viabahn® endoprosthesis with heparin bioactive surface.The vessel was pre-dilated prior to the advancement of the device.The device was advance and deployed in its intended location with no reported issue.Resistance was noted as the delivery catheter was being removed from the patient.As the delivery catheter was removed from the guidewire it was noted that the distal shaft was not attached.The distal shaft was separated from the delivery catheter and was located in the patients common iliac artery.The physicians suspected cause of the distal shaft separation was excessive calcification in the vessel.Another gore device was implanted to pin the distal shaft to the vessel wall.The patient tolerated the procedure.
 
Additional Manufacturer Narrative
Added h6 evaluation codes.Engineering evaluation: the dual lumen, hub, and transition components were returned to gore for evaluation.Evaluation of returned device confirms the reported distal shaft separation, but the cause for the distal shaft separation cannot be confirmed without the distal shaft.The transition bond is intact with no evidence of the distal shaft remaining in the transition.Updated investigation conclusion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® VIABAHN® ENDOPROSTHESIS WITH HEPARIN BIOACTIVE SURFACE
Common Device Name
STENT, SUPERFICIAL FEMORAL ARTERY
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL ECHO RIDGE B/P
3250 w. kiltie lane
flagstaff AZ 86005
Manufacturer Contact
marcos ayala
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key13675883
Report Number2017233-2022-02786
Device Sequence Number17734963
Product Code NIP
UDI-Device Identifier00733132622764
UDI-Public00733132622764
Combination Product (Y/N)Y
Initial Reporter StateKS
Initial Reporter CountryUS
PMA/510(K) Number
P040037
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/12/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/29/2024
Device Model NumberVBHR060502A
Device Catalogue NumberVBHR060502A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/07/2022
Supplement Date Received by Manufacturer02/07/2022
06/02/2026
Initial Report FDA Received Date03/04/2022
Supplement Report FDA Received Date06/03/2022
06/12/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-