| Model Number |
1012448-12 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/11/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a moderately calcified, moderately tortuous, 99% stenosed, de novo lesion in the distal left anterior descending (dlad) artery.Pre-dilatation was performed, and a non-abbott stent was deployed.Post-deployment, a 2.75x 12mm nc trek balloon dilation catheter (bdc) was advanced to the target lesion and inflated one time to 8 to 10 atmospheres (atm) when the balloon ruptured.The bdc was removed without issue.There was no adverse patient effect and no clinically significant delay in the procedure.A non-abbott bdc was used to successfully complete the procedure.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation determined the reported difficulty appears to be related to circumstances of the procedure.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
|
| |
|