Upon receipt at our post market quality assurance laboratory, detailed mechanical and electrical testing was performed on the device.The device battery status was end of life (eol).Elective replacement indicator (eri) was set prior to explant.The device functioned normally throughout testing.Laboratory analysis determined that this device experienced normal battery depletion; however, it was reported from the field that the estimated longevity remaining value appeared to decrease more quickly than expected between routine follow-ups.Factors influencing the estimated longevity remaining calculation include pacing rate, amplitude, pulse-width and lead impedance.Any (even slight) changes in these factors will impact the battery consumption calculation and therefore the remaining longevity estimate.Please note that, despite the drop in estimated longevity remaining, the actual battery condition did not change significantly between follow-ups.In summary, it was determined that this device experienced normal battery depletion, but declared eol earlier than previously estimated.Of note, the estimated remaining battery longevity of this pacemaker was designed to be calculated (and trigger corresponding device replacement indicators) based on the pacing capacitor charge time, which progressively increases as the battery depletes.These pacemakers were not designed with an ability to calculate remaining longevity for every possible current drain, which may cause replacement indicators to trigger earlier than what was previously estimated; however, it does not negatively impact the use, operation, safety, or performance of the device.
|
|
It was reported that this implantable device was suspected to be exhibiting premature battery depletion behavior.During the most recent interrogation, it was observed that the device had reached end of life (eol).The previous device interrogation showed that the remaining battery longevity had 2.5 years remaining.Reprogramming changes were made, and the outputs were increased.A copy of device memory was saved and submitted to boston scientific technical services (ts) for analysis.Engineering review, confirmed normal device operation, and discussed that the increase in output settings has the potential to pull elective replacement indicator (eri) forward.This device had been implanted for nearly 9.5 years.This device was explanted for normal eri.No adverse patient effects were reported.
|