Model Number S404
Medical Device Problem Code Premature Discharge of Battery (1057)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/25/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory, detailed mechanical and electrical testing was performed on the device.The device battery status was end of life (eol).Elective replacement indicator (eri) was set prior to explant.The device functioned normally throughout testing.Laboratory analysis determined that this device experienced normal battery depletion; however, it was reported from the field that the estimated longevity remaining value appeared to decrease more quickly than expected between routine follow-ups.Factors influencing the estimated longevity remaining calculation include pacing rate, amplitude, pulse-width and lead impedance.Any (even slight) changes in these factors will impact the battery consumption calculation and therefore the remaining longevity estimate.Please note that, despite the drop in estimated longevity remaining, the actual battery condition did not change significantly between follow-ups.In summary, it was determined that this device experienced normal battery depletion, but declared eol earlier than previously estimated.Of note, the estimated remaining battery longevity of this pacemaker was designed to be calculated (and trigger corresponding device replacement indicators) based on the pacing capacitor charge time, which progressively increases as the battery depletes.These pacemakers were not designed with an ability to calculate remaining longevity for every possible current drain, which may cause replacement indicators to trigger earlier than what was previously estimated; however, it does not negatively impact the use, operation, safety, or performance of the device.
 
Event or Problem Description
It was reported that this implantable device was suspected to be exhibiting premature battery depletion behavior.During the most recent interrogation, it was observed that the device had reached end of life (eol).The previous device interrogation showed that the remaining battery longevity had 2.5 years remaining.Reprogramming changes were made, and the outputs were increased.A copy of device memory was saved and submitted to boston scientific technical services (ts) for analysis.Engineering review, confirmed normal device operation, and discussed that the increase in output settings has the potential to pull elective replacement indicator (eri) forward.This device had been implanted for nearly 9.5 years.This device was explanted for normal eri.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALTRUA 40
Common Device Name
IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key13677544
Report Number2124215-2022-03111
Device Sequence Number5466216
Product Code LWP
UDI-Device Identifier00802526484452
UDI-Public00802526484452
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
N970003/S106
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date12/06/2012
Device Model NumberS404
Device Catalogue NumberS404
Device Lot Number597228
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/15/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/01/2022
Initial Report FDA Received Date03/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/08/2010
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-