• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION DYNAGEN MINI ICD VR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION DYNAGEN MINI ICD VR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT) Back to Search Results
Model Number D021
Medical Device Problem Code Defective Device (2588)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/11/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory, this device was thoroughly inspected and analyzed.Review of device memory identified an error.The error resulted in software resets performed in an attempt to correct an identified memory inconsistency.The cause of the memory inconsistency was not able to be determined through laboratory testing but was likely the result of exposure to radiation, either therapeutic or environmental.
 
Event or Problem Description
It was reported that this implantable cardioverter defibrillator (icd) was returned from our final pack area for analysis as the product had not passed all required testing.This icd was not implanted, and no patient involvement was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DYNAGEN MINI ICD VR
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key13677557
Report Number2124215-2022-00687
Device Sequence Number12129090
Product Code LWS
UDI-Device Identifier00802526533884
UDI-Public00802526533884
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P960040/S306
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 03/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date08/24/2020
Device Model NumberD021
Device Catalogue NumberD021
Device Lot Number109049
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/11/2021
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/13/2022
Initial Report FDA Received Date03/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/05/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-