Model Number AU00T0
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/05/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device with the lens was returned loose in the opened carton.The plunger lock and lens stop have been removed.Viscoelastic was observed in the device.The plunger was retracted to mid-nozzle.The lens appeared to have been replaced into the device loading area on top of the plunger.One haptic was folded onto the anterior optic surface and dried in viscoelastic.The other haptic distal tip has become broken in the closing of the door.There was no damage observed to the device.Product history records were reviewed and documentation indicated the product met release criteria.A qualified viscoelastic was indicated.The root cause cannot be determined for the reported complaint of "lens stuck to incision".The lens appeared to have been replaced into the device loading area on top of the plunger for return.The plunger of the used device was retracted to mid-nozzle.The position of the plunger relative to the lens during delivery cannot be determined.The ifu(instructions to follow) instructs: after the lens has been advanced to the nozzle line, the lens should be visually inspected to determine the position of the haptics.The plunger should be in contact with the trailing optic edge.After confirming the lens is properly positioned and the haptics are folded properly, proceed with lens implantation.Proceeding with implantation of a misfolded haptic or a lens that appears to be ¿out of position¿ can result in a broken haptic or other negative outcome.No device damage was observed.A qualified viscoelastic was indicated.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A non healthcare professional reported that during implantation of an intraocular lens (iol) the lens stuck to incision/wound.There was patient contact.Additional information was received stating the patient was not hospitalized for the event and there was no patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ SINGLEPIECE IOL WITH ULTRASERT DELIVERY SYSTEM
Common Device Name
LENS, GUIDE, INTRAOCULAR
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key13677804
Report Number1119421-2022-00406
Device Sequence Number4894073
Product Code KYB
UDI-Device Identifier00380652394918
UDI-Public00380652394918
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 03/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/19/2024
Device Model NumberAU00T0
Device Lot Number15187565
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/21/2022
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/14/2022
Initial Report FDA Received Date03/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/21/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PROVISC
Patient SexUnknown
-
-