• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. TPRLC 133 T1 PPS SO 10X140MM T1; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. TPRLC 133 T1 PPS SO 10X140MM T1; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Medical Device Problem Code Device Contaminated During Manufacture or Shipping (2969)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/09/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during warehouse inspection, debris was found in the sterile packaging.There was no patient involvement.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Additional Manufacturer Narrative
(b)(4).Report source (b)(6).The customer has indicated that the product is in process of being returned to zimmer biomet for the investigation.Once the investigation is complete, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001825034 - 2022 - 00469, 0001825034 - 2022 - 00471, 0001825034 - 2022 - 00472.
 
Event or Problem Description
No additional event information to report at this time.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected updated: d4; d9; g3; h2; h3; h4; h6 visual inspection of the returned product identified debris inside the sterile package.Review of the device history records identified no related deviations or anomalies.These products likely left zimmer biomet control non-conforming.The root cause of the reported event can be attributed to the operator not following instructions during manufacturing.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TPRLC 133 T1 PPS SO 10X140MM T1
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key13678837
Report Number0001825034-2022-00470
Device Sequence Number5474642
Product Code LZO
UDI-Device Identifier00880304489684
UDI-Public(01)00880304489684(17)310528(10)7027117
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K200196
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 05/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number51-103100
Device Lot Number7027117
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/09/2022
Supplement Date Received by Manufacturer05/23/2022
Initial Report FDA Received Date03/07/2022
Supplement Report FDA Received Date05/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/28/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient SexUnknown
-
-