U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT GMBH TOTAL BILIRUBIN; DIAZO COLORIMETRY, BILIRUBIN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT GMBH TOTAL BILIRUBIN; DIAZO COLORIMETRY, BILIRUBIN Back to Search Results
Model Number 6L45-42
Medical Device Problem Code Low Test Results (2458)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/18/2022
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported a falsely decreased total bilirubin result generated on the architect c4000 analyzer on a patient.Results provided: (b)(6) 2022 (b)(6) =0.92 mg/dl, another architect = 1.25 mg/dl (normal range: 0.2-1.20 mg/dl).No impact to patient management was reported.
 
Additional Manufacturer Narrative
An evaluation is in process.A follow up report will be submitted when the evaluation is complete.
 
Additional Manufacturer Narrative
A review of tickets determined that there is normal complaint activity for lot 60072uq07.Trending review determined no trends for falsely depressed results for the product.Return testing was not completed as returns were not available.Manufacturing documentation for the likely cause lot did not identify any issues associated with the complaint issue.A review of the product labeling concluded that the issue is sufficiently addressed.Based on the investigation, no deficiency of the architect total bilirubin reagent lot number 60072uq07 was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOTAL BILIRUBIN
Common Device Name
DIAZO COLORIMETRY, BILIRUBIN
Manufacturer (Section D)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM  65205
Manufacturer (Section G)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM   65205
Manufacturer Contact
siobhan wright
lisnamuck
post market surveillance
longford N39 E-932
EI   N39 E932
433331157
MDR Report Key13681482
Report Number3002809144-2022-00073
Device Sequence Number14796404
Product Code CIG
UDI-Device Identifier00380740174972
UDI-Public00380740174972
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K121985
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2022
Device Model Number6L45-42
Device Catalogue Number06L45-42
Device Lot Number60072UQ07
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/18/2022
Supplement Date Received by Manufacturer03/24/2022
Initial Report FDA Received Date03/07/2022
Supplement Report FDA Received Date03/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/26/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARC C4000 INTGR, 02P24-40, C400758; ARC C4000 INTGR, 02P24-40, C400758
Patient SexUnknown
-
-