| Model Number |
1581-15-112 |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Hematoma (1884); Pain (1994); Joint Laxity (4526); Unspecified Musculoskeletal problem (4535)
|
| Date of Event |
02/18/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Patient underwent a right medial unicompartmental knee replacement many years ago (not a depuy implant).A number of years later the patient presented with a broken tibial component.At this time in 2015, the unicompartmental knee replacement was revised to a primary pfc femoral component with a stemmed tibial component (depuy cement was not utilised).The medial side of the tibia was bone grafted with an autograft and held in place with two k wires.Approximately 3 years after this "revision" the patient reports having been hit/ rammed by a cow and experienced some form of trauma around the medial aspect of the right knee.Patient has had progressively increasing pain, very specifically around the medial epicondyle of the femur.A decision was made to open the knee and investigate the cause and to look to change the bearing.On opening the knee, the synovial fluid was found to be blood stained.There also appeared to be traces of blackened metalosis (this has been attributed to the original uni knee which had broken as no metal on metal articulation was evident) a boney spur was evident around the medial epicondyle and there appeared to be wear to the poly bearing in the areas where this "spur" may have come into contact with the poly.The knee appeared to be a little lax in flexion vs extension.The boney spur was removed, the wound washed out and a slightly thicker insert implanted (from a 12.5mm cr curved to a 15mm cr curved plus insert) the knee appeared stable.Doi: unk 1st dor: 2015, 2nd dor: feb 18, 2022, right knee.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: device associated with this report was not received for examination.All available x-rays and photographs were reviewed, and no evidence of implant fracture, disassociation, or anything indicative of a device nonconformance was found.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed, and the investigation may be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|