Model Number FIBERTAG TIGHTROPE WITH FLIPCUTTER III
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/11/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event or Problem Description
On 02/14/2022, it was reported by a sales representative via sems that an ar-1288rtt-fc3 fibertag tightrope with flipcutter iii had an issue.During an acl reconstruction on (b)(6) 2022, when drilling the femoral socket, the tip of the flipcutter popped off when the cutting adjustable end hit bone.The broken piece was retrieved from the patient.Continued the case with a new flipcutter.The procedure was completed successfully with no harm to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FIBERTAG TIGHTROPE WITH FLIPCUTTER III
Common Device Name
PIN, FIXATION, SMOOTH
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key13684450
Report Number1220246-2022-04554
Device Sequence Number4124455
Product Code HTY
UDI-Device Identifier00888867357624
UDI-Public00888867357624
Combination Product (Y/N)N
Initial Reporter StateNV
Initial Reporter CountryUS
PMA/510(K) Number
K112990
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 03/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2025
Device Model NumberFIBERTAG TIGHTROPE WITH FLIPCUTTER III
Device Catalogue NumberAR-1288RTT-FC3
Device Lot Number13008620
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/14/2022
Initial Report FDA Received Date03/07/2022
Date Device Manufactured03/29/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-